5.5 RECOGNITION CHARACTER AND START AND STOP CHARACTERS
The meter recognizes an asterisk (
*
) as the command recognition character. In the counter,
another command recognition character may be chosen to make the meter compatible with an
existing system. The meter will still respond to an asterisk. For all meters, in continuous
mode, a device ,such as a printer, may require a start and stop bit to recognize the data string
being sent. Normally there is no start bit and the stop bit is a carriage return <cr>. When the
Counter is in a normal operating mode (not Remote Display), SER 3, XDXXX can be set for
the following combinations:
Command Continuous Readings
D Recognition Char Start Char Stop Char
0
*
None CR
1 Selected None CR
2
*
Selected Selected
3 Selected Selected Selected
Either Instrument Setup.exe or Serial.exe may be used to set the Command recognition char-
acter and the start stop characters. These programs may be downloaded from our website.
6.0 MEMORY ADDRESSES AND DATA DEFINITIONS
6.1 DPM
1-BYTE RAM DATA TABLE
Hex Hex
Address Item Name Value
E1 Configuration Bit 7 6 5 4 3 2 1 0
0 = Linear Data
1 =Custom Curve (Extended DPM)
0 = Full Duplex (RS485)
1 = Half Duplex
0 = Un-Latched RTS
1 = Latched RTS
0 = Setup scale method
1 = Reading 2 coord method
0 0 0 = Not Rate
0 0 1 = Rate X 0.1
0 1 0 = Rate X 1
0 1 1 = Rate X 10
1 0 0 = Rate X 100
1 0 1 = Rate X 1000
1 1 0 = Rate X 10,000
0 = Not used
Vedere la pagina 21